Declaring the End of a Cooldown

A Shield Transient Biases the Flatness Criterion in an Anti-Stokes Optical Refrigerator

A paper from the School of Settled Matters and the School of Continuous Improvement holds 41 laser cooldowns of one ytterbium-doped fluoride sample for ten hours each and finds that a passively mounted clamshell, warming over about 144 minutes, turns the late approach into a shallow dip, so that the laboratory's ten-minute flatness criterion reports a median 1.00 K colder than the sample's equilibrium in every run. A sequential rule that refits a two-time-constant thermal model and waits for its asymptote to hold still comes within 0.5 K in all 41 runs, for a median 59 more minutes of pumping.